Item
|
Specifications
|
| Titration process |
| Standard: |
Potentiometric titration |
| Optional: |
Photometric titration, Volumetric Karl-Fischer titration |
|
| Measuring
Range |
pH: 0 to 14, Resolution
0.01pH
mV: -2000 to +2000 mV, Resolution 0.1 mV |
| Electrode
Inputs |
Indicator electrode
: 1
Reference electrode : 1
Temperature electrode : 1
KF electrode : 1 |
| Titration
steps |
Up to three (3)
steps |
| Number
of Endpoints |
Up to two (2) points |
| End point
Detection |
Automatic detection
by:
-Inflection point
-Set point
-F Cross point
-B Cross point |
| Display
screen
|
2-lines by 16digits with backlit LCD |
| Display
unit |
pH, mV, mgH2O,
%, ppm, M, g/L, mg/g, Fact1, Fact2 |
| pH
Calibration |
2 or
3 points (compliant to ASTM or other standard
buffers) |
| File
Function |
30
files for potentiometric titrations
3 files for Karl Fischer titrations
10 files for titration control methods |
| Calculation
function |
Concentration
calculation,
, Statistics
calculation |
| Memory
data capacity |
50 titration results
along with titration curves and
parameters and via external computer |
Precautions
against
power failure |
Built-in backup
battery |
| Electrodes (Standard component) |
GE-101B Glass electrode, RE-201 Reference electrode, TE-401 Thermistor electrode |
| Buret |
Capacity: 20 mL, Minimum drop size: 0.01mL
Reproducibility: within +/-0.01 mL |
| Tubing |
PTFE tube of 2 mm(ID) and 3 mm(OD) |
| Switching Valve |
Automatic switching valve with ceramic joint. |
| Stirrer |
Integrated, variable
speed, magnetic stirrer |
| Printers |
Optional thermal printer; PR-2000T3,
Optional dot impact printer; PR-302B(CBM-910-40PF) |
| External
Input/Output |
| RS-232C; 2 ports |
| Port 1 (BALANCE): |
For connection to electronic
balance. |
| Port 2 (HOST): |
For connection to external
computer. |
|
| Dimensions |
310mm (W) x 270mm
(D) x 310mm (H); |
| Weight |
Approx.7 kg |
| Power
Supply
; Power consumption |
AC 100 - 240 V, 50/60 Hz & 100 VA |