FEATURES
-
1. Six (6) built-in calculation modes accommodate various states of samples
-
2. Four (4) files of preset conditions can be stored in memory
-
3. Free software for downloading titration results to a P.C.; includes the RS-232C cable
-
4. CD training movie and data acquisition software“AQUANet”are included
-
5. Simple structure of electrolytic cell for easy maintenance
SPECIFICATIONS
Titration method | Coulometric Karl Fischer Titration |
---|---|
Electrolysis control method | Constant-current electrolysis, intermittent electrolysis near the end point |
Titration control method | AC polarization potentiometric detection method |
Measuring range | 10μgH₂O – 99mgH₂O |
Accuracy | Within CV of 0.3%
(1g of 1mgH₂O/mL water-in-methanol measured 10 times) |
Required measuring time | FAST: 1.2mgH₂O / min. (20μgH₂O / sec.)
SLOW: 0.6mgH₂O / min. (10μgH₂O / sec.) |
Electrolytic cell capacity | 150mL |
Electrolysis diaphragm | Ion exchange membrane |
Display | 16-Digit x 2-Line LCD with backlighting |
Display unit | μgH₂O, %, ppm |
File function | 4 files for titration parameters |
Data storage capacity | 20 data |
External input/Output | 2 RS-232C ports for Balance and Host, 1 interface for printer |
Software | Data acquisition software is included. |
GLP compliant functions | Maintenance information on reagents, electrode, membrane |
Power requirement | AC100-240V, 50/60Hz, 100VA |
Dimensions / Weight | 310(W) x 270(D) x 170(H) mm / Approx. 5kg |