Solid Evaporator EV-2000

Simple measurement of moisture
in solid and powdered samples.
Vaporizers used to efficiently measure the moisture in solid and powdered samples such as various resin pellets, medicines (powdered and solid), various inorganic powders and granules, copy machine toner, etc.
*Since sales availability, products and services vary by country and region, please contact your Hitachi High-Tech Analysis contact, the exclusive distributor, for specific requests and questions.
*If the message "This video cannot be played" is displayed, please try Super Reload.
[Windows:Ctrl+F5 / Mac:Command+Shift+R
Product features
1. Heating temperature (room temperature to 350℃)
Efficiently vaporizes the moisture in solid and powdered samples.
(*Samples which generate moisture through decomposition by heat such as sodium bicarbonate and samples which emit substances that react with Karl Fischer reagent cannot be measured.)
2.Simple setting of measurement conditions
Using an interface connection with the moisture measurement device it is easy to set conditions for the sample with a measurement sequence settings file installed in the moisture measurement device (including heating temperature conditions).
※ MOICO-A19/MOIVO-A19/AQ-2200A/AQV-2200A
Specification
Temperature control | Optional settings, automatic temperature adjustment |
---|---|
Temperature range | From room temperature to 350 ℃ |
Carrier gas | N2 gas (or dry air from an optional dry air pump) Flow rate : 0.1 to 1.0L / min (depending on setting) |
Carrier gas Desiccant | Molecular sieves 100g |
Heater | Band heater |
Sample amount | About 10mL or less |
Power supply | AC 100-240V 50/60Hz About 150VA A power cable is not included in this product. (*1) |
Dimensions / Weight | 220(W) x 430(D) x 200(H) mm / Approx. 8kg |
1: The power inlet of this product complies with IEC60320 C14.
Please prepare a power cable that can be connected to this power inlet.
NOTICE: Cannot be used in combination with single chamber electrolytic cell for AQ-300 or AQ-7.
Option
- dry air pump
- Optional key-pad
*required when the EV-2000 is used with AQ/AQV-300