Automated Oil Evaporator LE-320

Continuous automatic measurement system
for lubricating oil for 20 samples
Equipped with automatic reagent
exchange function
Create a system by combining it with a KF Titrator.
It can continuously and efficiently measure lubricating oil, grease, paint, etc., which are difficult to measure by directly injecting them into the cell.
*Since sales availability, products and services vary by country and region, please contact your Hitachi High-Tech Analysis contact, the exclusive distributor, for specific requests and questions.
Product features
1.Automatic measurement
Multiple samples can be measured automatically for efficient moisture measurement.
2. Heating temperature (room temperature to 200℃)
The sample heating temperature can be set arbitrarily from room temperature to 200°C.
3.Measurement condition file
There are 50 measurement condition files, and optimal temperature settings can be registered for each sample.
4.Temperature increase can be set for each sample
Up to 3 steps of temperature increase can be set for each sample.
5.Easy-to-handle sample bottles
Since PTFE is used for the packing, there is no blank from the packing.
6.Automatic reagent exchange function is standard
The reagent replacement can be set by the number of measurements to automatically drain and replace the generator electrolyte.
Specification
For KF titrator, please refer to the specifications of the KF titrator to be combined.
Moisture vaporization method | Azeotropic Distillation Method |
---|---|
Sample container | Hard glass Bottle, aluminum foil packing, PTFE lid |
Number of sample containers | 20 samples |
Sample amount (volume) | About 3 mL or less |
Toluene Dispensing Volume | About 5 mL (per measurement) |
Carrier gas | N2 gas Flow rate : 30 to 300 mL / min (depending on setting) |
Reagent exchange | Automatic exchange by setting the number of measurements |
Carrier gas Desiccant | Molecular sieves |
Heater/Temperature control | Band heater, PID control method |
Temperature range | From room temperature to 200 ℃ |
End position setting | When the stop pin is set to the end position or no container is detected |
Post-measurement procedures | End buzzer |
Power supply | AC 100 - 240 V 50/60 Hz About 150 VA A power cable is not included in this product. (*1) |
Dimensions / Weight | Sample changer: 350(W)x 510(D) x 390(H) mm / approx. 19kg Valve box: 275(W) x 440(D) x 280(H) mm / approx. 9kg |
*1: The power inlet of this product complies with IEC60320 C14.
Please prepare a power cable that can be connected to this power inlet.